Bump Test Gas — Complete Guide for Gas Detector Safety Programs

Bump Test Gas — Complete Guide for Gas Detector Safety Programs

Bump testing is one of the most important — and most commonly neglected — procedures for portable gas detector programs. This guide covers what bump test gas is, how often to do it, which gas to use for your detector, and how to set up an automatic supply program so you never run out. The gas we supply is produced in ISO/IEC 17025:2017-accredited laboratories, ships with a NIST-traceable CoA, and is available in all standard mixtures and sizes.

What is a bump test?

A bump test (function check) briefly exposes a detector to a known concentration of calibration gas to verify that each sensor responds and that the detector triggers its alarms. It does not adjust the sensor readings — it only confirms the detector responds. A detector that fails a bump test must be removed from service for full calibration or repair.

How often should you bump test?

Standard / requirement Frequency
OSHA 29 CFR 1910.146 (confined space) Before each use
OSHA 29 CFR 1910.120 (HAZWOPER) Before each use
MSA / BW / Industrial Scientific manuals Before each use
ANSI/ISEA 104-2016 (IDLH) Before each use

Bottom line: for confined space entry, HAZWOPER, or any IDLH environment, bump test before every single use.

Bump test vs. calibration

Bump test Full calibration
Purpose Verify sensor responds and alarms Adjust sensor to known concentration
Duration 30–60 seconds 2–5 minutes per sensor
Adjusts detector? No Yes
Frequency Before each use Every 30–180 days
Gas used per event ~0.1–0.2 L ~0.5–1 L

Which bump test gas do I need?

Detector configuration Recommended gas Part #
CO + H₂S + O₂ + LEL (standard) 25ppm H₂S / 100ppm CO / 50% LEL / 18% O₂ 34L-380
MSA Altair 4X 25ppm H₂S / 100ppm CO / 1.45% CH₄ / 15% O₂ 34L-421 / 58L-421
CO only 100 ppm CO / N₂ 34L-CO-100
H₂S only 25 ppm H₂S / N₂ 34L-H2S-25
O₂ only 18% O₂ / N₂ 34L-O2-18
NO₂ sensor 400 ppm NO₂ / N₂ 34L-NO2-400
SO₂ sensor 20 ppm SO₂ / N₂ 34L-SO2-20
NH₃ sensor 25 ppm NH₃ / Air 34L-NH3-25
Cl₂ sensor 10 ppm Cl₂ / N₂ 34L-CL2-10

Set up an Auto-Replenish program

Running out of bump test gas creates a compliance gap. To size a subscription: count your detectors × bump test frequency, at ~0.1–0.2 L per test, divided by cylinder volume. Rough guide: 17L for 1–2 detectors, 34L for 3–5, 58L for 6–10, 103L for 10+. Then choose a 30/60/90-day interval at trigas-usa.com for 8% off plus free shipping, with a reminder before each delivery and CoA/SDS included.

Bump test documentation — what to record

For confined space and HAZWOPER programs, record: date/time, instrument serial and model, calibration gas cylinder lot and part number, pass/fail per sensor, who performed the test, and the action taken on failure. Every Trigas USA cylinder includes a CoA whose lot number links back to the ISO 17025-accredited laboratory that produced the gas — a complete, auditable traceability chain to NIST standards.