4-Gas Calibration Cylinder — Complete Guide for CO, H₂S, O₂ and LEL Bump Testing

4-Gas Calibration Cylinder — Complete Guide for CO, H₂S, O₂, and LEL Bump Testing

The 4-gas calibration cylinder — commonly called a quad gas cylinder — is the most widely used calibration gas in portable gas detection. It contains four gases at once: Carbon Monoxide (CO), Hydrogen Sulfide (H₂S), Oxygen (O₂), and a combustible gas for Lower Explosive Limit (LEL) calibration. The gas we supply is produced in ISO/IEC 17025:2017-accredited laboratories, ships with a NIST-traceable Certificate of Analysis, and is available in all standard sizes for same-day shipping.

What is a 4-gas calibration cylinder?

Gas Symbol Standard concentration Hazard
Carbon Monoxide CO 100 ppm Toxic
Hydrogen Sulfide H₂S 25 ppm Toxic
Oxygen O₂ 18% (deficiency) Oxygen
Methane (LEL) CH₄ 50% LEL (2.5% CH₄) Combustible

The most common 4-gas mixture — the “380 mixture” — contains exactly these four gases at these concentrations and is the universal standard for most portable 4-gas detectors.

Bump test vs. full calibration

Bump test (function check): exposes the detector to gas to confirm each sensor responds and alarms — it does not adjust the detector. Recommended before each use, or at least daily for IDLH environments.

Full calibration (span): adjusts the sensor readings to match the known concentration, correcting drift. Typically every 30–180 days depending on use and sensor type.

Which cylinder size do I need?

Cylinder Approx. bump tests Approx. calibrations Best for
17L-380 ~100 ~30 1–2 detectors, field use
34L-380 ~200 ~60 Standard — most common
58L-380 ~350 ~100 Multiple detectors
103L-380 ~600 ~180 Safety teams, 5+ detectors

Common 4-gas mixtures — part numbers

Part # CO H₂S LEL / CH₄ O₂ Common detector
34L-380 100ppm 25ppm 50% LEL 18% MSA, BW, iSC
34L-383 100ppm 25ppm 2.5% CH₄ 18% Multiple brands
34L-421 100ppm 25ppm 1.45% CH₄ 15% MSA Altair 4X
34L-486 50ppm 10ppm 50% LEL 18% Low-range detectors
34L-491 50ppm 25ppm 0.7% Pentane 18% Pentane LEL sensor
58L-421 100ppm 25ppm 1.45% CH₄ 15% MSA Altair 4X fleet

Which mixture for my detector?

Detector Recommended mixture Part #
MSA Altair 4X 25ppm H₂S / 100ppm CO / 1.45% CH₄ / 15% O₂ 34L-421 or 58L-421
MSA Altair 5X 25ppm H₂S / 100ppm CO / 50% LEL / 18% O₂ 34L-380 or 58L-380
BW GasAlert Quattro 25ppm H₂S / 100ppm CO / 50% LEL / 18% O₂ 34L-380
BW Micro Clip XL 25ppm H₂S / 100ppm CO / 50% LEL / 18% O₂ 17L-380 or 34L-380
Industrial Scientific MX6 25ppm H₂S / 100ppm CO / 50% LEL / 18% O₂ 34L-380 or 103L-380
Draeger X-am 5000 25ppm H₂S / 100ppm CO / 50% LEL / 18% O₂ 34L-380

Always verify the required concentration in your detector’s user manual — manufacturer specifications take precedence. Need help confirming? Email sales@trigas-usa.com.

Auto-Replenish for 4-gas cylinders

A 4-gas cylinder is a consumable. With Auto-Replenish you choose your mixture, size, and a 30/60/90-day interval for 8% off plus free shipping, with a reminder before each delivery and CoA/SDS included. Modify, pause, or cancel anytime; Net-30 available for B2B.

FAQ

What does 50% LEL mean? Half the Lower Explosive Limit. For methane (LEL 5% by volume), 50% LEL = 2.5% CH₄ — safe to ship while still triggering the combustible sensor.

Can I use any 4-gas cylinder with my detector? The mixture must match your sensor configuration and concentration range. Most use the 380 mixture; some (e.g. MSA Altair 4X) need 15% O₂. Check your manual.

How long does a 34L-380 last? Roughly 150–200 bump tests, depending on regulator flow, test duration, and temperature.